Feature
without SPC data output
Order No. | 293-241-30 |
---|---|
Range (mm) | 25 - 50 |
Resolution (mm) | 0.001 |
Measuring force*1 (N) | 5 - 10 |
Maximum permissible error J MPE (µm) | ±1 |
Parallelism (µm) | 1 |
Flatness (µm) | 0.3 |
Constant measuring force | With ratchet stop |
Mass (g) | 330 |
Digital/Analog: | Digital |
Inch-Metric: | Metric |
• Dust /water protection level: IP65 (IEC60529)*2
• Battery:SR44 (1 pc.), 938882, for initial operational checks (standard accessory)
• Battery life: Approx. 2.4 years under normal use
• Length standard: Electromagnetic rotary sensor
• Standard accessories: Setting standard, 1 pc. (except for 0 to 25 mm (0 to 1 in) models), Spanner (301336), 1 pc.
*1 Measuring force when using the speeder ratchet (Apply a measuring force in the same condition as for measurement and then set the origin.)
*2 Rustproofing shall be applied after use.
Note: All digits of models over 125 mm (5 in) measuring range are presettable.