CNC Rockwell Hardness Testing Machine
HR-610A/620A
The HR-600 Series can measure and test large, heavy objects without cutting. Simply load and test.
Testing can also be automated with an electrically powered X/Y table loader.
Further automation can be implemented by linking with a transport or a signal tower.
Test force: 29.42 - 1,839 N (3 - 187.5 kgf), 9.807 - 2,452 N (1 - 250 kgf)
Maximum load: 100 kg
Applicable standards and test force
HR-610A | HR-620A | ||
Test methods/ Standard No. |
Rockwell | JIS B7726:2017, ISO 6508-2:2015, ASTM E18-20 | |
Brinell* | JIS B7724:2017, ISO 6506-2:2017, ASTM E10-18 | ||
Plastic | - | ISO 2039-1:2001 | |
JIS K7202-2:2001, ISO 2039-2:1987, ASTM D785-08 [A&B] | |||
Indentation Brinell hardness | VDI/VDE 2616 | ||
Indentation Vickers hardness | - | VDI/VDE 2616 | |
Initial test force N (kgf) |
Rockwell | 29.42 (3) 98.07 (10) | |
Plastic | - | 9.807 (1) | |
98.07 (10) | |||
Indentation Brinell hardness | 98.07 (10) 490.3 (50) | ||
Indentation Vickers hardness | - | 9.807 (1) | |
Test force N (kgf) |
Rockwell | 147.1 (15) 294.2 (30) 441.3 (45) 588.4 (60) 980.7 (100) 1471 (150) | |
Brinell | 49.03 (5) - 1839 (187.5) | 9.807 (1) - 2452 (250) | |
Plastic | 49.03 (5) 132.4 (13.5) 358.0 (36.5) 962.1 (98.1) | ||
588.4 (60) 980.7 (100) 1471 (150) | |||
Indentation Brinell hardness | 612.9 (62.5) 1839 (187.5) | 612.9 (62.5) 1839 (187.5) 2452 (250) | |
Indentation Vickers hardness | - | 294.2 (30) 490.4 (50) |
* For Brinell hardness testing, an indenter (optional) and a measurement microscope are required. A measurement microscope should be prepared by customer.
Specifications
Model | HR-610A | HR-620A | ||||
Unit (display unit) | metric | inch/mm | metric | inch/mm | ||
Indenter type*¹ | 1/16" Steel ball | 1/16" Tungsten carbide ball | 1/16" Steel ball | 1/16" Tungsten carbide ball | ||
Testable hardness | Rockwell hardness/Rockwell Superficial hardness/Brinell hardness/ Indentation Brinell hardness/Plastics hardness |
|||||
Test force range | 29.42 to 1839 N (3 to 187.5 kgf) | 9.807 to 2452 N (1 to 250 kgf) | ||||
Specimen height (Z-axis stroke) | 40 to 250 mm | |||||
Workpiece criteria |
Minimum surface dimensions | 18×4 mm or more | ||||
Minimum inner diameter of pipe-type workpiece |
ø400 mm or more | |||||
Concave workpiece | R25 mm or more, Height 20 mm or less | |||||
Minimum outside diameter | ø20 mm or more | |||||
Z-axis speed | Approx. 10 mm/s | |||||
Maximum depth (from indenter center) | 220 mm | |||||
X-axis stroke | None (Option: 160 mm or 300 mm) | |||||
Y-axis stroke | None | |||||
Maximum table loading | 100 kg | |||||
Display | Standard | Number of data displayed: 1, Hardness value, Scale, Test numbering, Hold time (Initial test force), Hold time (Total test force), Reading time, Hardness conversion, Judgment, Correction, Unit |
||||
Simple | Number of data displayed: 1, Hardness value, Scale, Judgment, Correction | |||||
List average/list | Number of data displayed: 5, Hardness value, Average hardness value, Hardness variation, Scale, Hardness conversion, Judgment, Correction | |||||
Multipoint | Depending on set test points, Hardness value, Scale, Test numbering, Judgment, Correction, Unit | |||||
Calculation | GO/NG judgment function | Test results judged according to set maximum/minimum values | ||||
Conversion function | Converts obtained test results to another scale | |||||
Correction functions |
Curved surface compensation | Corrects results according to specimen shape (cylindrical, spherical, etc.) | ||||
User correction |
Shift | Corrects by increasing/decreasing value according to hardness value | ||||
Multipoint | Corrects based on the results of tests conducted on multiple standard blocks (Rockwell/superficial only) | |||||
External output settings |
Serial | For printer (RS-232C compliant) 1-ch | ||||
Digimatic | Digimatic interface outputs 1-ch | |||||
USB2.0 | For USB memory/1-ch for PC communication | |||||
Languages | Supports the following 15 languages Japanese, English, German, French, Italian, Spanish, Korean, Chinese (simplified/traditional), Turkish, Portuguese, Polish, Czech, Hungarian and Dutch |
|||||
Hardness value | Digital display | Max. 7-digit (including decimal point and sign) | ||||
Minimum reading | 0.01 (settings can be changed) | |||||
Average hardness valu | Average value of valid data | |||||
Hardness variation | Variations in valid data (Max. - Min. ) | |||||
Scale | HRC/HR15N/HBW2.5/187.5 etc. | |||||
Display | Hardness value, test condition, OK/NG judgment result, statistical calculation result, X-R control chart, hardness conversion value | |||||
Test numbering | When testing a single specimen: 1, 2, 3... | |||||
When testing a group of specimens: 1/5-1, 2/5-1, 3/5-1, 4/5-1, 5/5-1, 1/5-2, 2/5-2... | ||||||
Hold time | Initial test force | 1 to 120 s (configurable in seconds) | ||||
Total test force | 1 to 120 s (configurable in seconds) | |||||
Reading time | 0 to 120 s (configurable in seconds) | |||||
Hardness conversion | MITUTOYO HARD STEEL, SOFT METAL/SAE J417 T1/ASTM E140 T1, T2, T4 ISO 18265 TA. 1/BS 860 T2, T3, T4 | |||||
Judgment | OK, ±NG | |||||
Correction | Displays whether or not correction was applied: cylindrical, spherical, user (multipoint/shift correction) Unit mm |
|||||
Unit | mm (X-, Y- and Z-axis stage displacements) | |||||
Power supply | AC100 to 240 V 50/60 Hz | |||||
Mass | 176 kg | 181 kg |
*1 Supplied as standard.
Note 1: Plastic tests may not be supported depending on the plastic material.
Note 2: For Brinell hardness testing, an indenter (optional) and a measurement microscope are required. A measurement microscope should be prepared by customer.
Note 3: No indenter and hardness standard block is supplied with the unit. These items (conform to the applicable standard) must be purchased separately.
แบรนด์สินค้า | MITUTOYO |
---|